Philips HeartStart FRx Semi Automatic Defibrillator
The Philips FRx AED is designed for use by both trained and untrained responders. It ensures responders of all levels can use the defibrillator whilst attending to a medical emergency and feeling confident in the treatment they are delivering.
Features of the FRx AED:
-
The SMART Pads II can be used on both adults and children
-
Wireless Data Transfer for the transmission of information
-
The Infant / Child Key provides a signal to the device for the treatment of minors
-
Flashing icons indicate pad placement on the patient
-
A CPR coaching function that can be used on adults and children
-
SMART Analysis: technology for heart rhythm assessment
-
SMART Biphasic technology allows treatment to be modified for each specific patient
-
IP55 rating
HeartStart FRx
The Philips HeartStart FRx AED features step-by-step voice instructions, including CPR guidance, and an audible metronome to assist guide basic life support (BLS) while treating a suspected sudden cardiac arrest. The Pre-connected SMART Pads II can be used on both adults and children. The device is rugged, lightweight, and reliable and withstands extreme temperatures and rough handling. Philips HeartStart FRx is the partner by your side when every minute counts.
Features
Patented technology. Proven therapy.
Step-by-step, real-time voice commands which are paced to your actions with an audible metronome and CPR guidance assist the responder. Whilst treating an infant or child, simply insert the optional infant/child key and the FRx will adjust the therapy and instructions accordingly.
Designed to work where you need it
The Philips HeartStart FRx AED is lightweight, rugged, and reliable and withstands rough handling, extreme temperatures, and dusty and wet surfaces. It is designed for use in harsh settings and can withstand up to 500 kg (1,100 lbs) and drops from 1.2 m (4 ft).
Easy as 1-2-3
The Quick Shock feature allows the FRx to deliver a shock in 8 seconds after performing CPR. Studies show that minimizing the time to shock after performing CPR can improve survival rates. (1) Press the green On/Off button to activate a voice instruction with visual icons. (2) Place the pads on the patient. (3) Press the orange Shock button when advised.
Simplified Maintenance
The pre-connected SMART Pads II can be used on both adults and minors. Once installed and activated the FRx is simple to maintain. It conducts a series of automatic self-tests, daily, weekly, and monthly to ensure pad readiness with verified functionality and calibration of systems and circuits. It has a life span of up to four years between battery replacements.
Specifications
Product Specifications:
Model Number
861304
Waveform
-
Truncated Exponential Biphasic
-
Waveform parameters are adjusted as a function of each patient’s impedance
Therapy
-
Adult defibrillation: 32A (150J nominal) to a 50ohm load.
-
Paediatric defibrillation (with Infant/Child Key installed): 19A (50J nominal) to a 50ohm load.
Battery
4 years when maintained and stored according to the directions provided in the Owner’s Manual
Protocol
-
Protocol Device follows the pre-configured settings.
-
The defibrillation and CPR protocol can be customiSed using software of the HeartStart Event Review or HeartStart Configure
Quick Shock
Able to deliver a shock after the end of the CPR interval, usually in 8 seconds
Voice Instructions
-
Real-Time Operating Instructions means the instructions and voice prompts given by the AED keep pace with the pace of the responder
-
If the responder is quicker, the AED will follow.
-
If the responder is slower, the AED will wait, rephrase and repeat instructions.
-
AEDs which are given a red cross "play back" instructions, like a tape recorder at a constant speed, irrespective of the pace of the responder.
CPR Guidance
Instructions for adult and infant / child CPR available
Pacemaker Detection
Capable of detecting a pacemaker and adjusting the heart rhythm analyses and measurements accordingly
Environmental Specifications:
Temperature
Operating / Standby: 32° - 122° F / 0° - 50° C
Electromagnetic interference (EMI)
-
CISPR II Group I Class B, IEC 61000-4-3
-
IEC 61000-4-8
Sealing
-
Waterjet proof IPX5 per IEC60529
-
Dust protected IPX5 per IEC60529
Vibration
-
Operating: meets MILSTD 810F Fig.514.5C-17
-
Random Standby: meets MILSTD 810F Fig.514.5C-18, swept sine.